Cryogen-free dilution refrigerator – BF1
Model | LD250 |
---|---|
Manufacturer | Bluefors |
Internal designation | BF1 |
Localization | D2-0066 |
Technical resource(s) | Christian Lupien Michael Lacerte |
Types of analysis | Electrical measurements at very low temperatures. Possibility to apply a 7 T magnetic field. |
Quantitative, semi-quantitative, qualitative | Quantitative |
Sample size (mm) | min: - max: diameter 30 cm x 48 cm height |
Temperature (K) | min: 10 mK max: 1.5 K |
Magnetic field (T) | min: - 7T max: 7 T |
Measurements: DC or microwave? | DC and microwave measurements |
Time required to cool the sample down | Approximately 24 h (except when the magnet is installed) |
Specific features | Cryogen-free equipment - liquid helium not used Cooling power: 335 μW at 100 mK 4 high frequency lines (18 GHz), 24 DC electrical lines and 24 filtered thermocoax lines Possibility to add 4 K 40 GHz lines 1 optical fiber |
What makes it unique? | Possibility to install a 7 T magnet. |